Strany 1 - Instruction Manual

Instruction ManualCalibration FacilityCFR

Strany 2 - IMPORTANT USER INFORMATION

CFR manual 9 Step 5 Connect the red and black signal measurement cables to either the back or front input terminals of the voltmeter and to

Strany 3 - Manual version: 0107

CFR manual 10 2.3 Maintenance In general, the manual calibration facility requires little maintenance apart from keeping it cl

Strany 4 - Declaration of Conformity

CFR manual 11 “fresh” value. If the scale change is more than 1%, critical examination of the radiation centre procedures for old and new ref

Strany 5 - Table of contents

CFR manual 12 There are several procedures for transferring calibration values from a narrow field of view instrument (pyrhel

Strany 6 - Introduction

CFR manual 13 3 Calibration operation The indoor calibration procedure, according to ISO 9847 Appendix III, is based on a si

Strany 7 - Installation and operation

CFR manual 14 The possibility of making errors due to inhomogeneous light is still present. Therefore, the positions of the pyranometers are

Strany 8 - 2.2 Installation

CFR manual 15 3.2 Description of the procedure step-by-step 1. Switch on the lamp and the voltmeter. 2. Place the reference and test pyran

Strany 9

CFR manual 16 3.3 Calibration uncertainty The main purpose of the calibration procedure is to perform a one-to-one comparison o

Strany 10

CFR manual 17 3.3.3 Voltmeter offset Theoretically, the voltmeter can cause a zero offset or drift. Short-circuiting the voltme

Strany 11 - 2.3.1 Lamp

CFR manual 18 3.4 Calibration Form An example of a calibration form is given below. Test pyranometer model: CMP 6 Serial number: 060002

Strany 12

1 IMPORTANT USER INFORMATION Reading this entire manual is recommended for full understanding of the use of this product. Should

Strany 13

CFR manual 19 4 Principle components and specifications of the facility The Calibration Facility consists of a number of parts. These are

Strany 14 - 3 Calibration operation

CFR manual 20 4.3 Shading mechanism The shading mechanism serves to obtain the dark (zero-offset) signal of the pyranometers and the mea

Strany 15

CFR manual 21 4.7 Voltage Stabilization The voltage stabilizer provides a highly stable voltage to the lamp. For calibration it is necessary

Strany 16

CFR manual 22 The Kipp & Zonen reference pyranometers are calibrated at the WRC with the sun and sky as the radiation source. As a cons

Strany 17 - 3.3.2 Pyranometer offsets

CFR manual 23 5 Frequently asked questions □ Can I calibrate pyranometers from different manufacturers with the CFR? The CFR is a univers

Strany 18 - 3.3.3 Voltmeter offset

CFR manual 24 Appendix I: Calibration Form for pyranometers Test pyranometer model : ……………………….. Serial number : ……………………….. Ref

Strany 19

CFR manual 25 Appendix II: Pyranometer Classification According to WMO Guide 1996 Characteristics CMP 22 CMP 21 CMP 11 CMP 6 High qualit

Strany 20 - 4.1 Table

CFR manual 26 Appendix III: List of World and Regional Radiation Centres World Radiation Centres Davos (Switzerland) St. Petersburg (R

Strany 21

CFR manual 27 Appendix IV: Recalibration service Pyranometers, Albedometers, Pyrgeometers, UV-Radiometers & Sunshine Duration Sens

Strany 22 - 4.10 Reference pyranometer

Our customer support remains at your disposal for any maintenance or repair, calibration, supplies and spares.Für Servicearbeiten und Kalibrierung, Ve

Strany 23 - 4.12 Stopwatch

2 Manual version: 0107

Strany 24

3 Declaration of Conformity According to EC guideline 89/336/EEC We Kipp & Zonen B.V. Delftechpark 36 2628 XH D

Strany 25

CFR manual 4 Table of contents IMPORTANT USER INFORMATION ...

Strany 26

CFR manual 5 1 Introduction Dear customer, thank you for purchasing a Kipp & Zonen product. Please read this manual

Strany 27

CFR manual 6 2 Installation and operation 2.1 Delivery Check the contents of the shipment for completeness (see below) and note

Strany 28 - Sunshine Duration Sensors

CFR manual 7 2.2 Installation 2.2.1 Environmental conditions To perform accurate calibrations the calibration facility should prefer

Strany 29 - Passion for Precision

CFR manual 8 Step 3 Remove the mains plug from the lamp housing power cable (if fitted) and pass the cable through the lamp mounting pillar