Kippzonen POM-02 Sky Radiometer Uživatelský manuál Strana 21

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 21
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 20
- 20 -
・Azimuth angle and zenith angle is motor position.
・Unit of output data is ampere.
・Data is recorded as scattering angle order.
Without barometer installed, com is selected, data record 0000 / com is not selected, data record -999.
User Maintenance
Before maintenance, please end the operation program and disconnect power cable.
・Silica gel replacement
There is silica gel case under cable connector. Remove that and if color is changed pink, replace
with new one. When you set silica gel case back sensor, make sure not to loose.
・Lens cleaning
Remove black cover for both short wave and long wave tube with hex wrench.
Blow air on lens and squirt some liquid cleaner on the surface.
Hex wrench
Remove case with hand
Zobrazit stránku 20