94
Spectral selectivity ± 2%
percentage deviation of the
product of spectral absorptance
and spectral transmittance
from the corresponding mean
within 0.35 :m and 1.5 :m
Temperature response ± 1%
percentage deviation due to
change in ambient temperature
within an interval of -20 to +50 °C,
relative to 20 °C.
Tilt response 0.25%
percentage deviation from the
responsivity at 00 tilt (horizontal)
due to change in tilt from 0° to 90°
at 1000 W m irradiance
-2
Viewing angle 2 B sr
Irradiance 0 - 1400 W m (max.4000 W m )
-2 -2
Spectral range 305-2800 nm (50% points)
335-2200 nm (95% points)
Sensitivity between 7 and 25 :V W m
-1 2
Impedance 40-100 Ohm
Receiver paint Carbon black
Glass domes Schott K5 optical glass 2 mm thick, 30 mm and 50 mm outer
diameter
Desiccant Silicagel
Spirit level Sensitivity 0.1° (bubble half out of the ring) Coincide with base
of the instrument. Detector surface and base are coplanar within
0.1°
Materials Anodized aluminium case Stainless steel screws in stainless
steel bushes. White plastic screen of ASA
Drying cartridge PMMA
Weight 830 g
Cable length 10 m
Dimensions 91.5 mm total height, 150 mm diameter, 25 mm dome height,
50 mm dome diameter
B 2.4 Kipp and Zonen Delft BV PYRANOMETER CM 31 (additions/changes to CM 21)
Spectral Range 200-4000 nm (50% points)
290-3500 nm (95% points)
Spectral selectivity max. 2% in the spectral range 300 to 3000 nm
Zero off-set + 4 W m
-2
response to 200 W m
-2
Komentáře k této Příručce